Pulsating Vacuum Sterilizer Biomedical Waste Autoclave Application:
Pulsating Vacuum Sterilizer Biomedical Waste Autoclave with the features of PLC and HMI automatic program control, automatic temperature recording and printing, vacuum drying, etc. is suitable for hospitals, pharmaceutical factories, scientific research institutes and other units to sterilize and dry clothing, dressings, metal instruments, culture medium, etc.
Pulsating Vacuum Sterilizer Biomedical Waste Sterilizer Advantage:
1. Touch-Screen Operation
2. High Precision Pressure Controller
3. Printing Data And Curve
4. National Patent Gate Lock Structure

Pulsating Vacuum Sterilizer Biomedical Waste Sterilizer Structure:
1. Vessels is designed to the China National Standard-GB150(Steel Pressure Vessel)
2. National Patent door interlock device.
3. National Patent sealing technology closed inflatable seal ring with compressed air.
4. Control system is integrated by PLC and HMI, an advanced function for recording, printing, and storing.
5. The GMP validation port is provided for the insertion of 16 temperature sensors for validation.
6. Independent temperature monitoring system.
7. The steam condenser is fitted in the exhaust/vacuum line to condense the steam before entering inside the vacuum pump.
Open door style:

Lift gate-type sterilizer
The way of opening the door: Sliding up-and-down.
Pit type sterilizer
The way of opening the door: The door will first travel upwards a bit, and then open after turning the handle.
Pulsating Vacuum Sterilizer Biomedical Waste Sterilizer Working principle:

Pulsating Vacuum Sterilizer Biomedical Waste Autoclave Main Parameters and Utilities:
Design Pressure |
0.245Mpa |
Vaccum |
-0.095Mpa |
Working Pressure |
0.22Mpa |
Steam Pressure |
0.4-0.6Mpa |
Design Temperature |
139 |
Tap Water Pressure |
0.2-0.3Mpa |
Working Temperature |
134 |
Compressed Air Pressure |
0.6-0.8Mpa |
Heat Equilibrium |
≤±1℃ |
|
Leave a Reply